English
 
Help Privacy Policy Disclaimer
  Advanced SearchBrowse

Item

ITEM ACTIONSEXPORT

Released

Journal Article

Improvements of the Jena AMS system

MPS-Authors
/persons/resource/persons62571

Steinhof,  A.
Service Facility 14C Lab, Dr. A. Steinhof, Max Planck Institute for Biogeochemistry, Max Planck Society;

/persons/resource/persons62405

Hejja,  I.
Service Facility 14C Lab, Dr. A. Steinhof, Max Planck Institute for Biogeochemistry, Max Planck Society;

/persons/resource/persons62596

Wagner,  T.
Service Facility 14C Lab, Dr. A. Steinhof, Max Planck Institute for Biogeochemistry, Max Planck Society;

External Resource
No external resources are shared
Fulltext (restricted access)
There are currently no full texts shared for your IP range.
Fulltext (public)
There are no public fulltexts stored in PuRe
Supplementary Material (public)
There is no public supplementary material available
Citation

Steinhof, A., Hejja, I., & Wagner, T. (2010). Improvements of the Jena AMS system. Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 268(7-8), 902-905. doi:10.1016/j.nimb.2009.10.060.


Cite as: https://hdl.handle.net/11858/00-001M-0000-000E-DAFB-B
Abstract
A new slit stabilization circuit and a modified generator drive were installed on an AMS system based on a 3 MV Tandetron (TM), produced by High Voltage Engineering Europa (HVEE). Furthermore our test procedure used at the Jena AMS system for the rectifiers of the Cockroft-Walton generator is presented. (C) 2009 Published by Elsevier B.V.