A triple-resonant RF coil setup for 1H, 23Na and 39K MR imaging of the rat brain at 9.4T

MPS-Authors
/persons/resource/persons83787

Augath,  MA
Department Physiology of Cognitive Processes, Max Planck Institute for Biological Cybernetics, Max Planck Society;
Max Planck Institute for Biological Cybernetics, Max Planck Society;

Fulltext (public)
There are no public fulltexts stored in PuRe
Supplementary Material (public)
There is no public supplementary material available
Citation

Augath, M., Heiler, P., Kirsch, S., & Schad, L. (2009). A triple-resonant RF coil setup for 1H, 23Na and 39K MR imaging of the rat brain at 9.4T. Poster presented at 17th Annual Meeting of the International Society for Magnetic Resonance in Medicine (ISMRM 2009), Honolulu, HI, USA.


Cite as: http://hdl.handle.net/11858/00-001M-0000-0013-C551-7
Abstract
The measurement of the spatial distribution of pathologically changing ion concentrations of 23Na and 39K could offer a very promising approach in clinical diagnostics. For MR imaging of both elements and the acquisition of anatomical proton images in the same experiment without moving the subject or the RF coil we developped a triple-resonant RF coil setup for the rat head at 9.4T. Two perpendicular Helmholtz pairs for 23Na and 39K delivered a good homogeneity over the sample volume. The flat double-D 1H surface coil showed no signs of coupling with the other coils and good coverage for shimming and anatomical images.