Moment and direction of the spoiler gradient for effective artifact suppression in RF-spoiled gradient echo imaging

MPS-Authors
/persons/resource/persons84187

Scheffler,  K
Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society;

External Resource
No external resources are shared
Fulltext (restricted access)
There are currently no full texts shared for your IP range.
Fulltext (public)
There are no public fulltexts stored in PuRe
Supplementary Material (public)
There is no public supplementary material available
Citation

Leupold, J., Hennig, J., & Scheffler, K. (2008). Moment and direction of the spoiler gradient for effective artifact suppression in RF-spoiled gradient echo imaging. Magnetic Resonance in Medicine, 60(1), 119-127. doi:10.1002/mrm.21614.


Cite as: https://hdl.handle.net/11858/00-001M-0000-0013-C827-B
Abstract
Radiofrequency (RF) -spoiled gradient echo sequences were developed with the aim to produce images with T1 weighted contrast within short acquisition time. Over the past two decades, this type of sequence has proven to be a robust technique and represents a reliable workhorse in clinical MRI. This study presents an analysis of ghost artifacts, which appear occasionally in RF-spoiled gradient echo images. It is demonstrated that the artifacts result from intrinsically emerging signal oscillations, which can be damped down by the application of spoiler gradients.