Is TrueFISP a gradient-echo or a spin-echo sequence?

MPS-Authors
There are no MPG-Authors in the publication available
External Resource
Fulltext (restricted access)
There are currently no full texts shared for your IP range.
Fulltext (public)
There are no public fulltexts stored in PuRe
Supplementary Material (public)
There is no public supplementary material available
Citation

Scheffler, K., & Hennig, J. (2003). Is TrueFISP a gradient-echo or a spin-echo sequence? Magnetic Resonance in Medicine, 49(2), 395-397. doi:10.1002/mrm.10351.


Cite as: https://hdl.handle.net/11858/00-001M-0000-0013-DCF4-F
Abstract
It is commonly accepted that TrueFISP (balanced FFE, FIESTA) belongs to the class of gradient-echo (GRE) sequences. GRE sequences are sensitive to dephasing effects of the transverse magnetization between the excitation pulse and echo acquisition, and phase coherence is only established directly after and before excitation pulses. However, an analysis of the phase evolution of transverse magnetization in a TrueFISP experiment shows very close similarities to the echo formation of a spin-echo (SE) experiment. If dephasing between excitation pulses is below ±π, TrueFISP exhibits a nearly complete refocusing of transverse magnetization at TE = TR/2. Only signals acquired before and after TR/2 show an additional Tmath image sensitivity.