Easy improvement of signal-to-noise in RARE-sequences with low refocusing flip angles

MPS-Authors
/persons/resource/persons84187

Scheffler,  K
Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society;

External Resource
There are no locators available
Fulltext (restricted access)
There are currently no full texts shared for your IP range.
フルテキスト (公開)
公開されているフルテキストはありません
付随資料 (公開)
There is no public supplementary material available
引用

Hennig, J., & Scheffler, K. (2000). Easy improvement of signal-to-noise in RARE-sequences with low refocusing flip angles. Magnetic Resonance in Medicine, 44(6), 983-985. doi:10.1002/1522-2594(200012)44:6<983:AID-MRM23>3.0.CO;2-8.


引用: https://hdl.handle.net/11858/00-001M-0000-0013-E3DA-3
要旨
It is demonstrated that the signal intensity in a RARE (TSE, FSE…)-sequence with low refocusing flip angle α can be significantly increased by setting the flip angle of the first refocusing pulse to 90°+α/2. In addition to the gain in signal intensity, the initial signal modulations over the first few echoes are reduced compared to a CPMG-echo train with constant α.