Functional MRI in human subjects with gradient-echo and spin-echo EPI at 9.4 T

MPS-Authors
/persons/resource/persons83838

Budde,  J
Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society;
Max Planck Institute for Biological Cybernetics, Max Planck Society;

/persons/resource/persons84213

Shajan,  G
Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society;
Max Planck Institute for Biological Cybernetics, Max Planck Society;

/persons/resource/persons84187

Scheffler,  K
Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society;
Max Planck Institute for Biological Cybernetics, Max Planck Society;

/persons/resource/persons84145

Pohmann,  R
Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society;
Max Planck Institute for Biological Cybernetics, Max Planck Society;

External Resource
Fulltext (restricted access)
There are currently no full texts shared for your IP range.
Fulltext (public)
There are no public fulltexts stored in PuRe
Supplementary Material (public)
There is no public supplementary material available
Citation

Budde, J., Shajan, G., Zaitsev, M., Scheffler, K., & Pohmann, R. (2014). Functional MRI in human subjects with gradient-echo and spin-echo EPI at 9.4 T. Magnetic Resonance in Medicine, 71(1), 209-218. doi:10.1002/mrm.24656.


Cite as: https://hdl.handle.net/11858/00-001M-0000-0027-8073-E
Abstract
PURPOSE:
The increased signal-to-noise ratio and blood oxygen level dependent signal at ultra-high field can only help to boost the resolution in functional MRI studies if the spatial specificity of the activation signal is improved. At a field strength of 9.4 T, both gradient-echo and spin-echo based echo-planar imaging were implemented and applied to investigate the specificity of human functional MRI. A finger tapping paradigm was used to acquire functional MRI data with scan parameters similar to standard neuroscientific applications.
METHODS:
Spatial resolution, echo, and readout times were varied to determine their influence on the distribution of the blood oxygen level dependent signal. High-resolution co-localized images were used to classify the signal according to its origin in veins or tissue.
RESULTS:
High-quality activation maps were obtained with both sequences. Signal contributions from tissue were found to be smaller or slightly larger than from veins. Gradient-echo echo-planar imaging yielded lower ratios of micro-/macro-vascular signals of around 0.6 than spin-echo based functional MRI, where this ratio varied between 0.75 and 1.02, with higher values for larger echo and shorter readout time.
CONCLUSION:
This study demonstrates the feasibility of human functional MRI at 9.4 T with high spatial specificity. Although venous contributions could not be entirely suppressed, venous effects in spin-echo echo-planar imaging are significantly reduced compared with gradient-echo echo-planar imaging.