Fast and efficient free induction decay MR spectroscopic imaging of the human brain at 9.4 Tesla

MPG-Autoren
/persons/resource/persons84408

Chadzynski,  GL
Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society;

/persons/resource/persons192600

Bause,  J
Max Planck Institute for Biological Cybernetics, Max Planck Society;
Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society;

/persons/resource/persons84213

Shajan,  G
Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society;

/persons/resource/persons84145

Pohmann,  R
Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society;
Max Planck Institute for Biological Cybernetics, Max Planck Society;

/persons/resource/persons84187

Scheffler,  K
Max Planck Institute for Biological Cybernetics, Max Planck Society;
Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society;

/persons/resource/persons83898

Ehses,  P
Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society;

Externe Ressourcen

Link
(beliebiger Volltext)

Volltexte (frei zugänglich)
Es sind keine frei zugänglichen Volltexte verfügbar
Ergänzendes Material (frei zugänglich)
Es sind keine frei zugänglichen Ergänzenden Materialien verfügbar
Zitation

Chadzynski, G., Bause, J., Shajan, G., Pohmann, R., Scheffler, K., & Ehses, P. (2017). Fast and efficient free induction decay MR spectroscopic imaging of the human brain at 9.4 Tesla. Magnetic Resonance in Medicine, 78(4), 1281-1295. doi:10.1002/mrm.26539.


Zitierlink: http://hdl.handle.net/21.11116/0000-0000-C29C-5
Zusammenfassung
Purpose The purpose of this work was to develop a fast and efficient MRSI-FID acquisition scheme and test its performance in vivo. The aim was to find a trade-off between the minimal total acquisition time and signal-to-noise ratio of the acquired spectra. Methods Measurements were performed on a 9.4 Tesla system. Sequence optimization included redesign of water suppression, optimization of the sequence gradients, and improvement of the sampling efficiency by minimizing the read-out time. This resulted in an acquisition time of 2:47 and 22:13 minutes for 2D (TR = 57 ms; 3-mm in-plane resolution) and 3D MRSI (TR = 57 ms; 16 slices; 3-mm isotropic resolution), respectively. Results Despite strong T1 weighting and first-order phase problems, it was possible to obtain spectra of an acceptable quality. The average line width calculated for the tCr peak across the entire field of view was 26.9 ± 9.6 Hz for 2D and 30.0 ± 11.3 Hz for 3D MRSI. In 3D measurements, the percent fraction of voxels fitted with Cramer-Rao lower bounds below 10 was 53.3 ± 4.1, 63.4 ± 8.4, and 81.0 ± 2.9 for Glu, tCr, and tNAA, respectively. Conclusion Considering the typically long duration of high-resolution MRSI, the proposed technique may be of interest for clinical applications and/or studies that focus on following the biochemistry of dynamic processes.