English
 
Help Privacy Policy Disclaimer
  Advanced SearchBrowse

Item

ITEM ACTIONSEXPORT

Released

Journal Article

Controlling motion artefact levels in MR images by suspending data acquisition during periods of head motion

MPS-Authors
/persons/resource/persons19617

Draganski,  Bogdan
Département des Neurosciences Cliniques, Laboratoire de Recherche en Neuroimagerie (LREN), Centre hospitalier universitaire vaudois, Lausanne, Switzerland;
Department Neurology, MPI for Human Cognitive and Brain Sciences, Max Planck Society;

External Resource
No external resources are shared
Fulltext (restricted access)
There are currently no full texts shared for your IP range.
Fulltext (public)

Castella_2018.pdf
(Preprint), 2MB

Supplementary Material (public)
There is no public supplementary material available
Citation

Castella, R., Arn, L., Dupuis, E., Callaghan, M. F., Draganski, B., & Lutti, A. (2018). Controlling motion artefact levels in MR images by suspending data acquisition during periods of head motion. Magnetic Resonance in Medicine, 80(6), 2415-2426. doi:10.1002/mrm.27214.


Cite as: https://hdl.handle.net/21.11116/0000-0002-9296-F
Abstract
PURPOSE:

Head movements are a major source of MRI artefacts. Prospective motion correction techniques significantly improve data quality, but strong motion artefacts may remain in the data. We introduce a framework to suspend data acquisition during periods of head motion over a predefined threshold.
METHODS:

Data was acquired with prospective motion correction and an external optical tracking system. A predictor of motion impact was introduced that accounts for the amplitude of the signal acquired at the time of the motion. From this predictor, a threshold was defined to trigger the suspension of data acquisition during periods of motion. The framework was tested on 5 subjects, 2 motion behaviors, and 2 head coils (20 and 64 channels).
RESULTS:

The best improvements in data quality were obtained for a threshold value of 0, equivalent to suspending the acquisition based on head speed alone, at the cost of a long prolongation of scan time. For threshold values ∼3.5e-4 , image quality was largely preserved, and prolongation of scan time was minimal. Artefacts occasionally remained with the 64-channel head coil for all threshold values, seemingly due to head movement in the sharp sensitivity profile of this coil.
CONCLUSION:

The proposed suspension strategy is more efficient than relying on head speed alone. The threshold for suspension of data acquisition governs the tradeoff between image degradation due to motion and prolonged scan time, and can be tuned by the user according to the desired image quality and participant's tolerability.