English
 
Help Privacy Policy Disclaimer
  Advanced SearchBrowse

Item

ITEM ACTIONSEXPORT
 PreviousNext  

Released

Journal Article

A nitrite microsensor for profiling environmental biofilms

MPS-Authors
/persons/resource/persons210257

de Beer,  Dirk
Permanent Research Group Microsensor, Max Planck Institute for Marine Microbiology, Max Planck Society;

/persons/resource/persons210761

Schramm,  Andreas
Permanent Research Group Microsensor, Max Planck Institute for Marine Microbiology, Max Planck Society;

/persons/resource/persons210740

Santegoeds,  Cecilia M.
Permanent Research Group Microsensor, Max Planck Institute for Marine Microbiology, Max Planck Society;

/persons/resource/persons210554

Kühl,  Michael
Permanent Research Group Microsensor, Max Planck Institute for Marine Microbiology, Max Planck Society;

External Resource
No external resources are shared
Fulltext (restricted access)
There are currently no full texts shared for your IP range.
Fulltext (public)

De_Beer_1997.pdf
(Publisher version), 95KB

Supplementary Material (public)
There is no public supplementary material available
Citation

de Beer, D., Schramm, A., Santegoeds, C. M., & Kühl, M. (1997). A nitrite microsensor for profiling environmental biofilms. Applied and Environmental Microbiology, 63(3), 973-977.


Cite as: https://hdl.handle.net/21.11116/0000-0005-0A5D-9
Abstract
A highly selective liquid membrane nitrite microsensor based on the hydrophobic ion-carrier aquocyano-cobalt(III) -hepta(2-phenylethyl)-cobrynate is described, The sensor has a tip diameter of 10 to 15 mu m. The response is log-linear in freshwater down to 1 mu M NO2- and in seawater to 10 mu M NO2-. A method is described for preparation of relatively large polyvinyl chloride (PVC)-gelled liquid membrane microsensors with a tip diameter of 5 to 15 mu m, having a hydrophilic coating on the tip, The coating and increased tip diameter resulted in more sturdy sensors, with a lower detection limit and a more stable signal than uncoated nitrite sensors with a tip diameter of 1 to 3 mu m. The coating protects the sensor membrane from detrimental direct contact with biomass and can be used for all PVC-gelled liquid membrane sensors meant for profiling microbial mats, biofilms, and sediments. Thanks to these improvements, liquid membrane sensors can now be used in complex environmental samples and in situ, e.g., in operating bioreactors. Examples of measurements in denitrifying, nitrifying, and nitrifying/denitrifying biofilms from wastewater treatment plants are shown. In all of these biofilms high nitrite concentrations were found in narrow zones of less than 1 mm.