Astigmatism Corrected Common Path Probe for Optical Coherence Tomography

MPS-Authors
There are no MPG-Authors in the publication available
External Resource
No external resources are shared
Fulltext (restricted access)
There are currently no full texts shared for your IP range.
Fulltext (public)
There are no public fulltexts stored in PuRe
Supplementary Material (public)
There is no public supplementary material available
Citation

Singh, K., Yamada, D., & Tearney, G. (2017). Astigmatism Corrected Common Path Probe for Optical Coherence Tomography. LASERS IN SURGERY AND MEDICINE, 49(3 SI), 312-318. doi:10.1002/lsm.22554.


Cite as: https://hdl.handle.net/21.11116/0000-0004-B79F-B
Abstract
Background and Objectives: Optical coherence tomography (OCT) catheters for intraluminal imaging are subject to various artifacts due to reference-sample arm dispersion imbalances and sample arm beam astigmatism. The goal of this work was to develop a probe that minimizes such artifact.
Materials and Methods: Our probe was fabricated using a single mode fiber at the tip of which a glass spacer and graded index objective lens were spliced to achieve the desired focal distance. The signal was reflected using a curved reflector to correct for astigmatism caused by the thin, protective, transparent sheath that surrounds the optics. The probe design was optimized using Zemax, a commercially available optical design software. Common path interferometric operation was achieved using Fresnel reflection from the tip of the focusing graded index objective lens. The performance of the probe was tested using a custom designed spectrometer-based OCT system.
Results: The probe achieved an axial resolution of 15.6 mu m in air, a lateral resolution 33 mu m, and a sensitivity of 103 dB. A scattering tissue phantom was imaged to test the performance of the probe for astigmatism correction. Images of the phantom confirmed that this common-path, astigmatism-corrected OCT imaging probe had minimal artifacts in the axial, and lateral dimensions.
Conclusions: In this work, we developed an astigmatism-corrected, common path probe that minimizes artifacts associated with standard OCT probes. This design may be useful for OCT applications that require high axial and lateral resolutions. (C) 2016 Wiley Periodicals, Inc.