日本語
 
Help Privacy Policy ポリシー/免責事項
  詳細検索ブラウズ

アイテム詳細


公開

学術論文

Model-based myocardial T1 mapping with sparsity constraints using single-shot inversion-recovery radial FLASH cardiovascular magnetic resonance.

MPS-Authors
/persons/resource/persons15082

Frahm,  J.
Biomedical NMR Research GmbH, MPI for biophysical chemistry, Max Planck Society;

External Resource
There are no locators available
Fulltext (restricted access)
There are currently no full texts shared for your IP range.
フルテキスト (公開)

3165576.pdf
(出版社版), 4MB

付随資料 (公開)

3165576_Suppl_1.png
(付録資料), 101KB

3165576_Suppl_2.png
(付録資料), 150KB

3165576_Suppl_3.docx
(付録資料), 14KB

3165576_Suppl_4.avi
(付録資料), 31MB

引用

Wang, X., Kohler, F., Unterberg-Buchwald, C., Lotz, J., Frahm, J., & Uecker, M. (2019). Model-based myocardial T1 mapping with sparsity constraints using single-shot inversion-recovery radial FLASH cardiovascular magnetic resonance. Journal of Cardiovascular Magnetic Resonance, 21:. doi:10.1186/s12968-019-0570-3.


引用: https://hdl.handle.net/21.11116/0000-0004-B91E-B
要旨
BACKGROUND:

This study develops a model-based myocardial T1 mapping technique with sparsity constraints which employs a single-shot inversion-recovery (IR) radial fast low angle shot (FLASH) cardiovascular magnetic resonance (CMR) acquisition. The method should offer high resolution, accuracy, precision and reproducibility.
METHODS:

The proposed reconstruction estimates myocardial parameter maps directly from undersampled k-space which is continuously measured by IR radial FLASH with a 4 s breathhold and retrospectively sorted based on a cardiac trigger signal. Joint sparsity constraints are imposed on the parameter maps to further improve T1 precision. Validations involved studies of an experimental phantom and 8 healthy adult subjects.
RESULTS:

In comparison to an IR spin-echo reference method, phantom experiments with T1 values ranging from 300 to 1500 ms revealed good accuracy and precision at simulated heart rates between 40 and 100 bpm. In vivo T1 maps achieved better precision and qualitatively better preservation of image features for the proposed method than a real-time CMR approach followed by pixelwise fitting. Apart from good inter-observer reproducibility (0.6% of the mean), in vivo results confirmed good intra-subject reproducibility (1.05% of the mean for intra-scan and 1.17, 1.51% of the means for the two inter-scans, respectively) of the proposed method.
CONCLUSION:

Model-based reconstructions with sparsity constraints allow for single-shot myocardial T1 maps with high spatial resolution, accuracy, precision and reproducibility within a 4 s breathhold. Clinical trials are warranted.