GRAPPA Operator Shift Correction for Non-Cartesian Imaging Trajectories

MPS-Authors
There are no MPG-Authors in the publication available
Fulltext (restricted access)
There are currently no full texts shared for your IP range.
Fulltext (public)
There are no public fulltexts stored in PuRe
Supplementary Material (public)
There is no public supplementary material available
Citation

Deshmane, A., Seiberlich, N., Duerk, J., & Griswold, M. (2012). GRAPPA Operator Shift Correction for Non-Cartesian Imaging Trajectories. Poster presented at 20th Annual Meeting and Exhibition of the International Society for Magnetic Resonance in Medicine (ISMRM 2012), Melbourne, Australia.


Cite as: https://hdl.handle.net/21.11116/0000-0006-9ADE-3
Abstract
Gradient timing delays can cause deviations from the ideal trajectory in non-Cartesian imaging. The proposed method aims to correct for trajectory errors and determine the signal at the center of k-space. Paths between points acquired along mis-centered radial projections and the true center of k-space are traced by a gradient ascent algorithm which iteratively applies a GRAPPA operator to shift k-space data while continually evaluating an estimate of the echo peak signal. This method can be used to perform trajectory measurement without field monitoring or extra acquisitions, and signal at the true k-space center can be used for retrospective self-gated imaging.