Trace gas detection with cavity ring down spectroscopy

MPS-Authors
There are no MPG-Authors in the publication available
External Resource
No external resources are shared
Fulltext (restricted access)
There are currently no full texts shared for your IP range.
Fulltext (public)

1.1145562.pdf
(Publisher version), 2MB

Supplementary Material (public)
There is no public supplementary material available
Citation

Jongma, R. T., Boogaarts, M. G. H., Holleman, I., & Meijer, G. (1995). Trace gas detection with cavity ring down spectroscopy. Review of Scientific Instruments, 66(4), 2821-2828. doi:10.1063/1.1145562.


Cite as: https://hdl.handle.net/21.11116/0000-000C-082C-B
Abstract
Trace gas detection of small molecules has been performed with cavity ring down (CRD) absorption spectroscopy in the near UV part of the spectrum. The absolute concentration of the OH radical present in trace amounts in heated plain air due to thermal dissociation of H2O has been calibrated as a function of temperature in the 720–1125 °C range. Detection of NH3 at the 10 ppb level is demonstrated in calibrated NH3/air flows. Detection of the background Hg concentration in plain air is performed with a current detection limit below 1 ppt. The effect of the laser linewidth in relation to the width of the absorption line is discussed in detail. Basic considerations regarding the use of CRD for trace gas detection are given and it is concluded that CRD spectroscopy holds great promise for sensitive [(sub)‐ppb] and fast (kHz) detection of many small molecules.