Scanning SQUID microscopy in a cryogen-free dilution refrigerator

MPS-Authors
There are no MPG-Authors in the publication available
External Resource
Fulltext (restricted access)
There are currently no full texts shared for your IP range.
フルテキスト (公開)
公開されているフルテキストはありません
付随資料 (公開)
There is no public supplementary material available
引用

Low, D., Ferguson, G. M., Jarjour, A., Schaefer, B. T., Bachmann, M. D., Moll, P. J. W., & Nowack, K. C. (2021). Scanning SQUID microscopy in a cryogen-free dilution refrigerator. Review of Scientific Instruments, 92(8):. doi:10.1063/5.0047652.


引用: https://hdl.handle.net/21.11116/0000-000B-75D6-0
要旨
We report a scanning superconducting quantum interference device (SQUID) microscope in a cryogen-free dilution refrigerator with a base temperature at the sample stage of at least 30 mK. The microscope is rigidly mounted to the mixing chamber plate to optimize thermal anchoring of the sample. The microscope housing fits into the bore of a superconducting vector magnet, and our design accommodates a large number of wires connecting the sample and sensor. Through a combination of vibration isolation in the cryostat and a rigid microscope housing, we achieve relative vibrations between the SQUID and the sample that allow us to image with micrometer resolution over a 150 µm range while the sample stage temperature remains at base temperature. To demonstrate the capabilities of our system, we show images acquired simultaneously of the static magnetic field, magnetic susceptibility, and magnetic fields produced by a current above a superconducting micrometer-scale device.