English
 
Help Privacy Policy Disclaimer
  Advanced SearchBrowse

Item

ITEM ACTIONSEXPORT

Released

Poster

Silent slice-wise shimming with a multi-coil setup at ultrahigh-field MRI

MPS-Authors
/persons/resource/persons192852

Aghaeifar,  A       
Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society;

/persons/resource/persons84187

Scheffler,  K       
Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society;

Fulltext (restricted access)
There are currently no full texts shared for your IP range.
Fulltext (public)
There are no public fulltexts stored in PuRe
Supplementary Material (public)
There is no public supplementary material available
Citation

Aghaeifar, A., & Scheffler, K. (2023). Silent slice-wise shimming with a multi-coil setup at ultrahigh-field MRI. Poster presented at 2023 ISMRM & ISMRT Annual Meeting & Exhibition (ISMRM 2023), Toronto, Canada.


Cite as: https://hdl.handle.net/21.11116/0000-000D-3596-E
Abstract
Multi-Coil shimming is advantageous because of its ease of use for effective local and slice-wise shimming. Shim coils are constructed of current-carrying conductor wires and undergo Lorenz force while placed in the magnet. Currents alteration in slice-wise shimming changes the magnitude and direction of force and imposes mechanical vibrations and acoustic noise. Here we propose strategies to control the level of inter-slice shims current change. The proposed strategies showed that inter-slice shims current change can be appropriately reduced by applying a minor change into shimming routine and without degrading shimming performance significantly.