English
 
Help Privacy Policy Disclaimer
  Advanced SearchBrowse

Item

ITEM ACTIONSEXPORT

Released

Journal Article

In situ UV/Vis/near-IR diffuse reflection measurement of catalysts at temperatures up to 673 K

MPS-Authors
/persons/resource/persons22169

Thiede,  Manfred
Inorganic Chemistry, Fritz Haber Institute, Max Planck Society;

/persons/resource/persons21863

Melsheimer,  Jörg
Inorganic Chemistry, Fritz Haber Institute, Max Planck Society;

External Resource
No external resources are shared
Fulltext (restricted access)
There are currently no full texts shared for your IP range.
Fulltext (public)
There are no public fulltexts stored in PuRe
Supplementary Material (public)
There is no public supplementary material available
Citation

Thiede, M., & Melsheimer, J. (2002). In situ UV/Vis/near-IR diffuse reflection measurement of catalysts at temperatures up to 673 K. Review of Scientific Instruments, 73(2), 394-397. doi:10.1063/1.1430730.


Cite as: https://hdl.handle.net/11858/00-001M-0000-0011-162E-E
Abstract
In-situ UV/VIS/near-IR diffuse reflectance spectroscopy can be carried out on catalysts with a suitable microreactor at temperatures up to 673 K. The experimental difficulties in the application of such a reactor at higher temperatures could be solved as follows:
Integrating sphere and reactor are connected by a ceramic part of high reflectivity and low thermal conductivity, i.e. the integrating sphere is enlarged. The signal/noise ratio increases by a factor of 2.
The standard IR-detector was replaced by a temperature-stabilized version. In addition watercooling jackets were mounted at the integrating sphere. The signal/noise ratio in the near-IR region was thus improved by a factor of 2-3 with a very good reproducibility of measurements.
To overcome an alteration of catalyst spectra by thermal radiation, a series of reference spectra of quartz powder (SiO2) were recorded at several temperatures. These spectra are used to correct the catalyst spectra by dividing catalyst and SiO2 spectra.