Integrated active tracking detector for MRI-guided interventions

MPS-Authors
There are no MPG-Authors in the publication available
Fulltext (restricted access)
There are currently no full texts shared for your IP range.
フルテキスト (公開)
公開されているフルテキストはありません
付随資料 (公開)
There is no public supplementary material available
引用

Anders, J., Sangiorgio, P., Deligianni, X., Santini, F., Scheffler, K., & Boero, G. (2012). Integrated active tracking detector for MRI-guided interventions. Magnetic Resonance in Medicine, 67(1), 290-296. doi:10.1002/mrm.23112.


引用: https://hdl.handle.net/11858/00-001M-0000-0013-B85E-5
要旨
We present a fully integrated detector suitable for active tracking of interventional devices in MR-guided interventions. The single-chip microsystem consists of a detection coil, a tuning capacitor, an intermediate frequency downconversion receiver, and a phase-locked-loop-based frequency synthesizer. Thanks to the integrated mixer, the chip output stage delivers an analog frequency-downconverted NMR signal in the frequency range from 0 to 200 kHz. The microchip, realized in a standard complementary metal oxide semiconductor technology, has a size of 1 × 2 × 0.74 mm3 and operates at a frequency of 63 MHz (i.e., in 1.5 T clinical scanners). Tests in a standard clinical scanner demonstrate the compatibility of the complementary metal oxide semiconductor microchip with clinical MRI systems. Using a solid sample of cis-polyisoprene having a size of 1 × 1.9 × 0.8 mm3 as internal signal source, the detector achieves a three-dimensional isotropic spatial resolution of 0.15 mm in a measuring time of 100 ms.