日本語
 
Help Privacy Policy ポリシー/免責事項
  詳細検索ブラウズ

アイテム詳細


公開

会議抄録

SQUID based magnetic resonance imaging for the investigation of in situ and in vivo hyperpolarization techniques

MPS-Authors
/persons/resource/persons133443

Buckenmaier,  K
Max Planck Institute for Biological Cybernetics, Max Planck Society;
Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society;

/persons/resource/persons214707

Rudolph,  M
Max Planck Institute for Biological Cybernetics, Max Planck Society;
Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society;

/persons/resource/persons216006

Fehling,  P
Max Planck Institute for Biological Cybernetics, Max Planck Society;
Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society;

/persons/resource/persons84187

Scheffler,  K
Max Planck Institute for Biological Cybernetics, Max Planck Society;
Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society;

Fulltext (restricted access)
There are currently no full texts shared for your IP range.
フルテキスト (公開)
公開されているフルテキストはありません
付随資料 (公開)
There is no public supplementary material available
引用

Buckenmaier, K., Rudolph, M., Fehling, P., Back, C., Bernarding, J., Koelle, D., Kleiner, R., Scheffler, K., & Plaumann, M. (2018). SQUID based magnetic resonance imaging for the investigation of in situ and in vivo hyperpolarization techniques. In 13th Annual Meeting of the European Society for Molecular Imaging (EMIM 2018).


引用: https://hdl.handle.net/21.11116/0000-0001-7E45-5
要旨
Introduction
Ultralow-field (ULF) nuclear magnetic resonance (NMR) is a promising spectroscopy method allowing for, e.g., the simultaneous detection of multiple nuclei. To overcome the low signal-to-noise ratio that usually hampers a wider application, we present an alternative approach to prepolarized ULF NMR employing hyperpolarization techniques like signal amplification by reversible exchange (SABRE) or Overhauser dynamic nuclear polarization (ODNP). Both techniques allow continuous hyperpolarization of 1H as well as other MR-active nuclei.
Methods
To be able to measure 1H and 19F simultaneously, a superconducting quantum interference device (SQUID)-based ULF NMR/MRI detection unit was constructed (see fig. 1). Due to the very low intrinsic noise level, SQUIDs are superior to conventional Faraday detection coils at ultralow-fields. Additionally, the broad band characteristics of SQUIDs enable them to simultaneously detect the MR signal of different nuclei such as 13C, 19F or 1H. Since SQUIDs detect the MR signal directly, they are an ideal tool for a quantitative investigation of hyperpolarization techniques such as SABRE or ODNP.
Results/Discussion
Using SABRE we successfully hyperpolarized fluorinated pyridine derivatives and quantitatively characterized the dependency of the magnetization transfer reaction from parahydrogen, which bonds to an iridium complex as well as to the 1H and 19F nuclei of an exchangeable ligand, as a function of hyperpolarization time and magnetic field strength [1]. Spectra (see fig. 2) and images of the samples were acquired.
With ODNP we were able to measure the coupling constant of solutions containing free radicals. Enhancement factors of over 100 were reached in in situ experiments. First proof-of-principle ex vivo images of rats using ODNP enhanced, SQUID based ULF-MRI have been acquired successfully.
Conclusions
We successfully built a SQUID-based ULF NMR/MRI system to quantitatively investigate the hyperpolarization techniques SABRE and ODNP.