  DeepCEST 3T: Robust MRI parameter determination and uncertainty quantification with neural networks-application to CEST imaging of the human brain at 3

Glang, F., Deshmane, A., Prokudin, S., Martin, F., Herz, K., Lindig, T., Bender, B., Scheffler, K., & Zaiss, M. (2020). DeepCEST 3T: Robust MRI parameter determination and uncertainty quantification with neural networks-application to CEST imaging of the human brain at 3. Magnetic Resonance in Medicine, 84(1), 450-466. doi:10.1002/mrm.28117.

Item is

基本情報

表示: 非表示:
アイテムのパーマリンク: https://hdl.handle.net/21.11116/0000-0005-5D92-E 版のパーマリンク: https://hdl.handle.net/21.11116/0000-000F-425E-E
資料種別: 学術論文

ファイル

表示: ファイル

関連URL

表示:
非表示:
説明:
-
OA-Status:
Not specified

作成者

表示:
非表示:
 作成者:
Glang, F1, 著者           
Deshmane, A1, 著者           
Prokudin, S, 著者
Martin, F1, 著者           
Herz, K1, 著者           
Lindig, T2, 著者           
Bender, B, 著者           
Scheffler, K1, 著者           
Zaiss, M1, 著者           
所属:
1Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society, ou_1497796              
2Institutional Guests, Max Planck Institute for Biological Cybernetics, Max Planck Society, ou_3505519              

内容説明

表示:
非表示:
キーワード: -
 要旨: Purpose: Calculation of sophisticated MR contrasts often requires complex mathematical modeling. Data evaluation is computationally expensive, vulnerable to artifacts, and often sensitive to fit algorithm parameters. In this work, we investigate whether neural networks can provide not only fast model fitting results, but also a quality metric for the predicted values, so called uncertainty quantification, investigated here in the context of multi‐pool Lorentzian fitting of CEST MRI spectra at 3T.
Methods

A deep feed‐forward neural network including a probabilistic output layer allowing for uncertainty quantification was set up to take uncorrected CEST‐spectra as input and predict 3T Lorentzian parameters of a 4‐pool model (water, semisolid MT, amide CEST, NOE CEST), including the B0 inhomogeneity. Networks were trained on data from 3 subjects with and without data augmentation, and applied to untrained data from 1 additional subject and 1 brain tumor patient. Comparison to conventional Lorentzian fitting was performed on different perturbations of input data.
Results

The deepCEST 3T networks provided fast and accurate predictions of all Lorentzian parameters and were robust to input perturbations because of noise or B0 artifacts. The uncertainty quantification detected fluctuations in input data by increase of the uncertainty intervals. The method generalized to unseen brain tumor patient CEST data.
Conclusions

The deepCEST 3T neural network provides fast and robust estimation of CEST parameters, enabling online reconstruction of sophisticated CEST contrast images without the typical computational cost. Moreover, the uncertainty quantification indicates if the predictions are trustworthy, enabling confident interpretation of contrast changes.

資料詳細

表示:
非表示:
言語:
 日付: 2019-122020-07
 出版の状態: 出版
 ページ: -
 出版情報: -
 目次: -
 査読: -
 識別子(DOI, ISBNなど): DOI: 10.1002/mrm.28117
 学位: -

関連イベント

表示:

訴訟

表示:

Project information

表示:

出版物 1

表示:
非表示:
出版物名: Magnetic Resonance in Medicine
種別: 学術雑誌
 著者・編者:
所属:
出版社, 出版地: New York : Wiley-Liss
ページ: - 巻号: 84 (1) 通巻号: - 開始・終了ページ: 450 - 466 識別子(ISBN, ISSN, DOIなど): ISSN: 0740-3194
CoNE: https://pure.mpg.de/cone/journals/resource/954925538149