Isotope ratio infrared spectroscopy analysis of water samples without memory effects

MPS-Authors
/persons/resource/persons252249

De Graaf,  Stefan
Climate Geochemistry, Max Planck Institute for Chemistry, Max Planck Society;

/persons/resource/persons204189

Vonhof,  Hubert B.
Climate Geochemistry, Max Planck Institute for Chemistry, Max Planck Society;

/persons/resource/persons252251

Levy,  Elan J.
Climate Geochemistry, Max Planck Institute for Chemistry, Max Planck Society;

/persons/resource/persons258066

Markowska,  Monika
Climate Geochemistry, Max Planck Institute for Chemistry, Max Planck Society;

/persons/resource/persons187781

Haug,  Gerald H.
Climate Geochemistry, Max Planck Institute for Chemistry, Max Planck Society;

External Resource
There are no locators available
Fulltext (restricted access)
There are currently no full texts shared for your IP range.
フルテキスト (公開)
公開されているフルテキストはありません
付随資料 (公開)
There is no public supplementary material available
引用

De Graaf, S., Vonhof, H. B., Levy, E. J., Markowska, M., & Haug, G. H. (2021). Isotope ratio infrared spectroscopy analysis of water samples without memory effects. Rapid Communications in Mass Spectrometry, 35(8):. doi:10.1002/rcm.9055.


引用: https://hdl.handle.net/21.11116/0000-0008-1F17-D
要旨

Rationale

Since their introduction more than a decade ago, isotope ratio infrared spectroscopy (IRIS) systems have rapidly become the standard for oxygen (δ18O) and hydrogen (δ2H) isotope analysis of water samples. An important disadvantage of IRIS systems is the well‐documented sample‐to‐sample memory effect, which requires each sample to be analyzed multiple times before the desired accuracy is reached, lengthening analysis times and driving up the costs of analyses.
Methods

We present an adapted set‐up and calculation protocol for fully automated analysis of water samples using a Picarro L2140‐i cavity ring‐down spectroscopy instrument. The adaptation removes memory effects by use of a continuously moisturized nitrogen carrier gas. Water samples of 0.5 μL are measured on top of the water vapor background, after which isotope ratios are calculated by subtraction of the background from the sample peaks.
Results

With this new technique, single injections of water samples have internal precisions (1σ) below 0.05‰ for δ18O values and 0.1‰ for δ2H values, regardless of the isotope ratio of the previous sample. Precision is worse, however, when the isotope difference between the sample and background water is too large (i.e., exceeding approximately 9‰ for δ18O values and 70‰ for δ2H values). Isotope ratios show negligible drift across the four weeks within which the experiments were performed. The single‐injection 1σ precision for 17O excess (Δ′17O) determined with this method is 60 per meg.
Conclusions

Our experiments demonstrate that by removing sample‐to‐sample memory effects with a moisturized carrier gas, the time for measurement of δ18O and δ2H values using an IRIS system can be reduced markedly without compromising the analytical precision and accuracy. Thorough replication is needed to achieve sufficiently low uncertainties for Δ′17O.