日本語
 
Help Privacy Policy ポリシー/免責事項
  詳細検索ブラウズ

アイテム詳細

  Signal-to-noise ratio versus field strength for small surface coils

Pohmann, R., Avdievich, N., & Scheffler, K. (2024). Signal-to-noise ratio versus field strength for small surface coils. NMR in Biomedicine, Epub ahead. doi:10.1002/nbm.5168.

Item is

基本情報

表示: 非表示:
アイテムのパーマリンク: https://hdl.handle.net/21.11116/0000-000F-5037-9 版のパーマリンク: https://hdl.handle.net/21.11116/0000-000F-5038-8
資料種別: 学術論文

ファイル

表示: ファイル

関連URL

表示:

作成者

表示:
非表示:
 作成者:
Pohmann, R1, 著者                 
Avdievich, NI1, 著者                 
Scheffler, K1, 著者                 
所属:
1Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society, ou_1497796              

内容説明

表示:
非表示:
キーワード: -
 要旨: The increasing signal-to-noise ratio (SNR) is the main reason to use ultrahigh field MRI. Here, we investigate the dependence of the SNR on the magnetic field strength, especially for small animal applications, where small surface coils are used and coil noise cannot be ignored. Measurements were performed at five field strengths from 3 to 14.1 T, using 2.2-cm surface coils with an identical coil design for transmit and receive on two water samples with and without salt. SNR was measured in a series of spoiled gradient echo images with varying flip angle and corrected for saturation based on a series of flip angle and T1 measurements. Furthermore, the noise figure of the receive chain was determined and eliminated to remove instrument dependence. Finally, the coil sensitivity was determined based on the principle of reciprocity to obtain a measure for ultimate SNR. Before coil sensitivity correction, the SNR increase in nonconductive samples is highly supralinear with B0 1.6-2.7, depending on distance to the coil, while in the conductive sample, the growth is smaller, being around linear close to the surface coil and increasing up to a B0 2.0 dependence when moving away from the coil. After sensitivity correction, the SNR increase is independent of loading with B0 2.1. This study confirms the supralinear increase of SNR with increasing field strengths. Compared with most human measurements with larger coil sizes, smaller surface coils, as mainly used in animal studies, have a higher contribution of coil noise and thus a different behavior of SNR at high fields.

資料詳細

表示:
非表示:
言語:
 日付: 2024-05
 出版の状態: オンラインで出版済み
 ページ: -
 出版情報: -
 目次: -
 査読: -
 識別子(DOI, ISBNなど): DOI: 10.1002/nbm.5168
 学位: -

関連イベント

表示:

訴訟

表示:

Project information

表示:

出版物 1

表示:
非表示:
出版物名: NMR in Biomedicine
種別: 学術雑誌
 著者・編者:
所属:
出版社, 出版地: London : Heyden & Son
ページ: - 巻号: Epub ahead 通巻号: - 開始・終了ページ: - 識別子(ISBN, ISSN, DOIなど): ISSN: 0952-3480
CoNE: https://pure.mpg.de/cone/journals/resource/954925574973