English
 
Help Privacy Policy Disclaimer
  Advanced SearchBrowse

Item

ITEM ACTIONSEXPORT
  Modeling radio-frequency energy-induced heating due to the presence of transcranial electric stimulation setup at 3T

Kozlov, M., Horner, M., Kainz, W., Weiskopf, N., & Möller, H. E. (2020). Modeling radio-frequency energy-induced heating due to the presence of transcranial electric stimulation setup at 3T. Magnetic Resonance Materials in Physics, Biology and Medicine, 33, 793-807. doi:10.1007/s10334-020-00853-5.

Item is

Files

show Files
hide Files
:
Kozlov_2020.pdf (Publisher version), 2MB
Name:
Kozlov_2020.pdf
Description:
-
OA-Status:
Hybrid
Visibility:
Public
MIME-Type / Checksum:
application/pdf / [MD5]
Technical Metadata:
Copyright Date:
-
Copyright Info:
-

Locators

show

Creators

show
hide
 Creators:
Kozlov, Mikhail1, Author           
Horner, Marc2, Author
Kainz, Wolfgang3, Author
Weiskopf, Nikolaus4, 5, Author           
Möller, Harald E.4, Author           
Affiliations:
1Department Neurophysics (Weiskopf), MPI for Human Cognitive and Brain Sciences, Max Planck Society, ou_2205649              
2ANSYS, Inc., Evanston, IL, USA, ou_persistent22              
3Division of Biomedical Physics, Office of Science and Engineering Laboratories (OSEL), Silver Spring, MD, USA, ou_persistent22              
4Methods and Development Unit Nuclear Magnetic Resonance, MPI for Human Cognitive and Brain Sciences, Max Planck Society, ou_634558              
5Felix Bloch Institute for Solid State Physics, University of Leipzig, Germany, ou_persistent22              

Content

show
hide
Free keywords: Computational modeling; RF simulations; Tissue heating; Transcranial direct current stimulation; Finite element method (FEM)
 Abstract: Purpose

The purpose of the present study was to develop a numerical workflow for simulating temperature increase in a high-resolution human head and torso model positioned in a whole-body magnetic resonance imaging (MRI) radio-frequency (RF) coil in the presence of a transcranial electric stimulation (tES) setup.
Methods

A customized human head and torso model was developed from medical image data. Power deposition and temperature rise (ΔT) were evaluated with the model positioned in a whole-body birdcage RF coil in the presence of a tES setup. Multiphysics modeling at 3T (123.2 MHz) on unstructured meshes was based on RF circuit, 3D electromagnetic, and thermal co-simulations. ΔT was obtained for (1) a set of electrical and thermal properties assigned to the scalp region, (2) a set of electrical properties of the gel used to ensure proper electrical contact between the tES electrodes and the scalp, (3) a set of electrical conductivity values of skin tissue, (4) four gel patch shapes, and (5) three electrode shapes.
Results

Significant dependence of power deposition and ΔT on the skin’s electrical properties and electrode and gel patch geometries was observed. Differences in maximum ΔT (> 100%) and its location were observed when comparing the results from a model using realistic human tissue properties and one with an external container made of acrylic material. The electrical and thermal properties of the phantom container material also significantly (> 250%) impacted the ΔT results.
Conclusion

Simulation results predicted that the electrode and gel geometries, skin electrical conductivity, and position of the temperature sensors have a significant impact on the estimated temperature rise. Therefore, these factors must be considered for reliable assessment of ΔT in subjects undergoing an MRI examination in the presence of a tES setup.

Details

show
hide
Language(s): eng - English
 Dates: 2020-04-222020-02-082020-05-122020-05-272020-12
 Publication Status: Issued
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Type: -
 Identifiers: DOI: 10.1007/s10334-020-00853-5
Other: epub 2020
PMID: 32462558
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Magnetic Resonance Materials in Physics, Biology and Medicine
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: Amsterdam : No longer published by Elsevier
Pages: - Volume / Issue: 33 Sequence Number: - Start / End Page: 793 - 807 Identifier: ISSN: 0968-5243
CoNE: https://pure.mpg.de/cone/journals/resource/954926245532