FastPtx: a versatile toolbox for rapid, joint design of pTx RF and gradient pulses using Pytorch's autodifferentiation

MPG-Autoren
/persons/resource/persons215985

Bosch,  D       
Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society;

/persons/resource/persons84187

Scheffler,  K       
Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society;

Volltexte (beschränkter Zugriff)
Für Ihren IP-Bereich sind aktuell keine Volltexte freigegeben.
Volltexte (frei zugänglich)
Es sind keine frei zugänglichen Volltexte in PuRe verfügbar
Ergänzendes Material (frei zugänglich)
Es sind keine frei zugänglichen Ergänzenden Materialien verfügbar
Zitation

Bosch, D., & Scheffler, K. (2024). FastPtx: a versatile toolbox for rapid, joint design of pTx RF and gradient pulses using Pytorch's autodifferentiation. Magnetic Resonance Materials in Physics, Biology and Medicine, 37(1), 127-138. doi:10.1007/s10334-023-01134-7.


Zitierlink: https://hdl.handle.net/21.11116/0000-000E-0681-9
Zusammenfassung
Objective: With modern optimization methods, free optimization of parallel transmit pulses together with their gradient waveforms can be performed on-line within a short time. A toolbox which uses PyTorch's autodifferentiation for simultaneous optimization of RF and gradient waveforms is presented and its performance is evaluated.
Methods: MR measurements were performed on a 9.4T MRI scanner using a 3D saturated single-shot turboFlash sequence for [Formula: see text] mapping. RF pulse simulation and optimization were done using a Python toolbox and a dedicated server. An RF- and Gradient pulse design toolbox was developed, including a cost function to balance different metrics and respect hardware and regulatory limits. Pulse performance was evaluated in GRE and MPRAGE imaging. Pulses for non-selective and for slab-selective excitation were designed.
Results: Universal pulses for non-selective excitation reduced the flip angle error to an NRMSE of (12.3±1.7)% relative to the targeted flip angle in simulations, compared to (42.0±1.4)% in CP mode. The tailored pulses performed best, resulting in a narrow flip angle distribution with NRMSE of (8.2±1.0)%. The tailored pulses could be created in only 66 s, making it feasible to design them during an experiment. A 90° pulse was designed as preparation pulse for a satTFL sequence and achieved a NRMSE of 7.1%. We showed that both MPRAGE and GRE imaging benefited from the pTx pulses created with our toolbox.